What are the requirements for the installation and stability of the equipment?

The device must be placed on a stable surface of at least 500 mm x 500 mm with adequate lighting and ventilation. Although permanent anchoring is not required, the unit must have a secure footing to prevent tipping during operation.

E S

Michael E.

How to safely isolate the power and perform maintenance/cleaning?

Isolate the power by using the main switch at the rear or by unplugging the unit. For cleaning, use a damp cloth with mild detergent; the use of compressed air is strictly prohibited.

E S

Michael E.

What safety measures protect against moving parts and emergency situations?

The device features an interlocked lid that prevents operation when open and remains locked for 30 seconds after shutdown due to inertia. The main switch or emergency stop button cuts power immediately. Hearing and eye protection are required if operating without an acoustic hood.

E S

Michael E.

How is the control system designed and accessed?

The system is accessed via a front interface featuring a touchscreen and a backlit control knob. The interface is organized into logical zones for status, navigation, settings, and commands for intuitive use.

E S

Michael E.

What are the installation and stability requirements for the ZM 300?

The device must be placed on a stable surface of at least 500 mm x 500 mm with adequate lighting and ventilation. Operators must use hearing protection, safety goggles, and safety shoes during setup and use.

E S

Michael E.

How are moving parts and emergency stops handled?

The rotor is protected by an automatic cover lock that prevents operation while open. The main switch or emergency stop cuts power, but the rotor may continue to spin for approximately 30 seconds due to inertia.

E S

Michael E.

Which chemicals can be used for maintenance or grinding processes?

Sewing machine oil can be used for the closing piston, and mild detergents for cleaning. Liquid nitrogen or dry ice can be used for the pre-grinding embrittlement of samples.

E S

Michael E.

How is the control system accessed and operated?

The device is operated via a front interface featuring a touchscreen and a jog dial with visual guidance. The menu is structured logically into status, navigation, and setting zones using standardized symbols.